Command Descriptions
2-68
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
Table 2- 28: Vertical position ranges using a 1X probe
CH<x>:SCAle Position range
2mV/div ±1000 divs
5mV/div ±400 divs
10 mV/div ±200 divs
20 mV/div ±100 divs
50 mV/div ±40 divs
100 mV/div ±20 divs
200 mV/div ±10 divs
500 mV/div ±100 divs
1V/div ±50 divs
2V/div ±25 divs
5V/div ±10 divs
CH2:POSition 1.32E0
Positions the channel 2 input signal 1.32 divisions above the center
of the display.
CH1:POSition?
Might return –1.32E0, indicating that the current position of channel
1 is at –1.32 divisions.
CH<x>:PRObe
Sets or queries the attenuation factor of the specified channel or
voltage probes. The value of <x> can vary from 1 through 4 for
4-channel instruments or 1 through 2 for 2-channel instruments.
Vertical
Examples
Group