Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-69
CH<x>:PRObe{1|10|20|50|100|500|1000 }
CH<x>:PRObe?
1 sets the specified oscilloscope channel to 1X attenuation.
10 sets the specified oscilloscope channel to 10X attenuation.
20 sets the specified oscilloscope channel to 20X attenuation.
(TDS1000B, TDS2000B, and TPS2000 series only)
50 sets the specified oscilloscope channel to 50X attenuation.
(TDS1000B, TDS2000B, and TPS2000 series only)
100 sets the specified oscilloscope channel to 100X attenuation.
500 sets the specified oscilloscope channel to 500X attenuation.
(TDS1000B, TDS2000B, and TPS2000 series only)
1000 sets the specified oscilloscope channel to 1000X attenuation.
<NR1>
CH2:PRObe 1000
Sets channel 2 to 1000X attenuation.
CH1:PRObe?
Might return CH1:PROBE 10
CH<x>:SCAle
Sets or queries the vertical gain of the specified oscilloscope
channel. The value of <x> can vary from 1 through 4 for 4-channel
instruments or 1 through 2 for 2-channel instruments.
This command is equivalent to adjusting the front-panel VOLTS/
DIV knob.
Syntax
Arguments
Returns
Examples