Command Descriptions
2-70
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
Vertical
CH<x>:SCAle <NR3>
CH<x>:SCAle?
<NR3> is the gain, in volts or amps per division. For example, the
voltage range is 5 V/div to 2 mV/div when using a 1X volt age probe.
CH1:SCAle 100E–3
Sets the channel 1 gain to 100 mV/div.
CH2:SCAle?
Might return 1.0E0, indicating that the current V/div setting of
channel 2 is 1 V/div.
CH1:VOLts
CH<x>:VOLts
Sets or queries the vertical gain of the specified channel. The value
of <x> can vary from 1 through 4 for 4-channel instrum ents or 1
through 2 for 2-channel i nstruments.
This command is identical to the CH<x>:SCAle command and is
included for compatibility purposes. Only CH<x>:SCAle is returned
in response to a CH<x>? query.
Group
Syntax
Arguments
Examples
Related Commands