Section 9 - Diagnostics
9-12
VM700A Option 01 (NTSC) & Option 11 (PAL) User Manual
-- LOW LEVEL DEBUG MODE --
Key Selection (Press one of the indicated keys)
--- -------------------------------------------
(N) ... Run Instrument in -NORMAL- (Power Up) mode
(A) ... Run Instrument in -AUTO RESET- mode
(Q) ... Run Instrument in -QUICK START- mode
(T) ... Calibrate -TOUCHPANEL-
(D) ... Run -DEBUGGER-
(C) ... -CLEAR- FileSystem Nvram
(1) ... Do -CPU BOARD- Diagnostic
(2) ... Do -DISPLAY/FP BOARD- Diagnostic
(3) ... Do -MAIN EPROM/NVRAM BOARD- Diagnostic
(H) ... Run Instrument in -HIGH LEVEL DEBUG- mode
(=) ... Reprint Menu
***
Press on of the indicated keys ***
Figure 9-2: Debug mode terminal screen display
D - Run the Debugger: runs a troubleshooting aid. Intended for qualified
service personnel only.
C - Clear the Filesystem NVRAM: restores file system to original factory
condition.
WARNING. The Clear NVRAM Filesystem option deletes all user-created files:
functions, timed events, setup files, the works! Any files you may wish to use
later should be backed up to a PC disk (using remote operations capability of
the VM700A and a utility such as VMBACKIT) before this option is invoked.
When you choose the Clear NVRAM Filesystem option, the following
message appears on the computer/terminal screen:

Table of Contents

Other manuals for Tektronix VM700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix VM700A and is the answer not in the manual?

Tektronix VM700A Specifications

General IconGeneral
BrandTektronix
ModelVM700A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals