iTRANS 2
FIXED POINT SINGLE OR DUAL GAS MONITOR
WITH DUAL ANALOG OUTPUTS
USER MANUAL
6
77036429-EN
Revision 7.0
2.2 Sensor
Item
Descriptions
Sensor Housing
Material
Aluminum, Anodized, Explosion-proof: Class I, Divisions 1 and 2,
Groups B, C, D and Ex d IICT6 Gb (China) and Ex d IIB + H
2
T5 (IECEx)
Aluminum, Anodized w/Gore-Tex Membrane: Class I, Division 2,
Groups A, B, C, D and Ex nA IIC T5 (IECEx)
Dimensions 3.0 × 3.0 inches (76 × 76 mm)
Accuracy
< ± 3% Toxic and Oxygen
For Combustibles:
For test gas concentrations up to and including 50% of full scale, the
deviation shall not exceed ±3% of full scale gas concentration.
For test gas concentrations above 50% of full scale, the deviation shall
not exceed ±5% of full scale gas concentration.
Protection Class IP 66 or NEMA 4X
Table 2-1 Sensor Specifications
2.3 Display
The gas monitor has a 4-digit, 7-
segment LED display for each of 2 channels.
A dual-gas
sensor and sample
display are shown in Figure 2-2.
Figure 2-2 The Display (Dual-Gas
Monitor Shown)
2.4 Inputs Intrusive and Non-Intrusive
The gas monitor can be configured using intrusive and non-intrusive means. Both methods
of configuration are accomplished through physical inputs that are visible behind the glass panel
of the gas monitor.
A set of four keys are used when intrusive programming is appropriate (i.e., when the enclosure
can be removed and when the keys can be manually pressed). These keys are the mode,
increment (+), decrement (-), and enter keys. Refer to

Related product manuals