iTRANS 2
FIXED POINT SINGLE OR DUAL GAS MONITOR
WITH DUAL ANALOG OUTPUTS
USER MANUAL
77036429-EN
Revision 7.0
19
Figure 4-13 Setting the ModBus Address for a Stand-Alone Sensor
NOTE: If adding a second sensor to an existing module, set the ModBus address to ↑↑↑↑↓↓↓↓
which represents 11110000 binary (and 240 decimal). See 6 for more information on the
ModBus interface. (Note that DIP switches are pre-set at the factory for all dual-sensor units).
4.7 Wiring Conclusion
Once wiring is complete, place the electronics module back in the housing by pressing
the standoff banana jacks into the mating plugs. Be careful not to pinch any of the wiring. After
the module is in place, secure the windowed top back on the housing and power up the unit.