EasyManua.ls Logo

TELEDYNE OLDHAM SIMTRONICS iTRANS 2 - Page 20

TELEDYNE OLDHAM SIMTRONICS iTRANS 2
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iTRANS 2
FIXED POINT SINGLE OR DUAL GAS MONITOR
WITH DUAL ANALOG OUTPUTS
USER MANUAL
14
77036429-EN
Revision 7.0
Figure 4-3 Sensor Connector J3 on the
NOTE: For dual-sensor configurations, place both of the same colored wires in the appropriate
terminal block and firmly tighten.
NOTE: Use #18 AWG (0.9 mm²) shielded cable for remote sensors. Maximum distance is 200
meters.
NOTE: When wiring remote sensors to the , “485 B” on J3 should be connected to “B-
in the remote sensor enclosure, and “485 A” on J3 should be connected to “A+” in the remote
sensor enclosure.
NOTE: For remote or standalone sensors, there are four terminal blocks located in the remote
sensor housing. These terminal blocks are all tied together and follow the same wiring scheme
mentioned above.
Figure 4-4 Wiring Diagram for a Single On-board Sensor

Related product manuals