Model 300E Carbon Monoxide Analyzer
Instruction Manual
Operating Instructions
P/N 04288 REV: A 59
6.3.5. Time of Day Clock (CLK)
The M300E has a time of day clock that supports the AutoCal timer, time of day
TEST function, and time stamps on most COM port messages. To set the time-of-
day, press:
SETUP C.3
CFG DAS RNGE PASS
CLK
MORE EXIT
SETUP C.3
TIME-OF-DAY CLOCK
TIME DATE EXIT
SETUP C.3
CFG DAS RNGE PASS CLK MORE EXIT
Exit to Return
to the
Sample Menu
Enter Current
Date-of-Year
SETUP C.3 DATE: 01-JAN-02
0 1 JAN 0 2 ENTR EXIT
SETUP C.3 DATE:
01-JAN-02
0 1 JAN 0 2
ENTR
EXIT
SETUP C.3 TIME-OF-DAY CLOCK
TIME DATE
EXIT
Enter Current
Time-of-Day
SETUP C.3 TIME:
12:00
1 2 : 0 0
ENTR
EXIT
SETUP C.3 TIME: 12:00
1 2 : 0 0 ENTR EXIT
SAMPLE RANGE = 50.0 PPM CO =
< TST TST > CAL
SETUP