D Series Syringe Pumps
Section 1 Introduction
1-2
Table 1-1 30D Technical Specifications
CYLINDER CAPACITY 30.42 ml
POWER REQUIREMENTS
(Mains voltage line cord is a
“Disconnect Device”)
100 ± 10 VAC, 2.8 A maximum
117 ± 12 VAC, 2.4 A maximum
234 ± 23 VAC, 1.2 A maximum
LINE FREQUENCY 50 or 60 Hz
LINE VOLTAGE NOISE TOLERANCE 1.7
nominal rms line voltage, 10 µsecond pulses, any phase
angle, random or repetitive
DIMENSIONS PUMP CONTROLLER
Width: 27.18 cm 27.18 cm
Depth: 46.74 cm 30.48 cm
Height: 100 cm 13.59 cm
WEIGHT PUMP CONTROLLER
36.4 kg 2.96 kg
FLOW RATE RANGE 0.01 µl/min to 22 ml/min (for any pressure up to 2068 bar)
FLOW RATE ACCURACY
a
± 0.5% above 69 bar and above 0.25 ml/min
(maximum 1.0 µl/min seal leakage)
FLOW RATE DISPLAY RESOLUTION 0.01 µl/min (1.0 µl/min in Constant Pressure Mode)
ANALOG OUTPUT ACCURACY
b
± 0.2% 4-20 mA output, ± 0.5% voltage output using the 12 channel
output option
DISPLACEMENT RESOLUTION 1.825 nl /step
REFILL TIME 1.05 minutes
REFILL OR DEPRESSURIZATION RATE 0.01 µl/min to 29 ml/min at any pressure from 0 to 2068 bar
PRESSURE RANGE 0.6895 to 2068.4 bar
PRESSURE ACCURACY ± 0.2% of full scale at constant temperature
PRESSURE REPEATABILITY
c
± 0.05% of full scale within 48 hours at constant temperature
ZERO PRESSURE DRIFT ± 0.05% of full scale within 48 hours at constant temperature
PRESSURE DISPLAY RESOLUTION 6.895 kPa
AMBIENT TEMPERATURE RANGE 5 to 40°C
TEMPERATURE DRIFT ± 0.015% of full scale/°C
HUMIDITY 95% maximum
DEAD (HEADSPACE) VOLUME
d
0.88 ± 0.02 ml
POLLUTION DEGREE 2
INSTALLATION CATEGORY II
MAXIMUM ALTITUDE 2000 m
a. Using water at constant pressure and constant temperature above 69 bar and above 0.25 ml/min.
b. The analog output is an optional accessory.
c. Pressure repeatability specification is based on zeroing the pressure transducer before measurement. Refer to sub-section
ZERO PRESS in Section 3 of the manual for re-zeroing procedure.
d. Volume in and above the piston seal, head clearance at automatic shutoff, and inlet and outlet ports to the fittings.