D Series Syringe Pumps
Section 1 Introduction
1-6
Table 1-5 100DX Technical Specifications
CYLINDER CAPACITY 102.93 ml
POWER REQUIREMENTS
a
(Mains voltage line cord is a
“Disconnect Device”)
100 ± 10 VAC, 1.5 A maximum
117 ± 12 VAC, 1.5 A maximum
234 ± 23 VAC, 0.75 A maximum
LINE FREQUENCY 50 or 60 Hz
LINE VOLTAGE NOISE TOLERANCE 1.7
nominal rms line voltage, 10 µsecond pulses, any phase
angle, random or repetitive
DIMENSIONS PUMP CONTROLLER
Width: 27.18 cm 27.18 cm
Depth: 46.74 cm 30.48 cm
Height: 101.09 cm 13.59 cm
WEIGHT PUMP CONTROLLER
32.8 kg 2.96 kg
FLOW RATE RANGE 0.01 µl/min to 50 ml/min (for any pressure up to 689.5 bar)
FLOW RATE ACCURACY
b
± 0.3% (maximum 0.25 µl/min seal leakage)
FLOW RATE DISPLAY RESOLUTION 0.01 µl/min (1.0 µl/min in Constant Pressure Mode)
ANALOG OUTPUT ACCURACY
c
± 1% of selected range
DISPLACEMENT RESOLUTION 9.65 nl /step
REFILL TIME 1.72 minutes
REFILL OR DEPRESSURIZATION RATE 0.01 µl/min to 60 ml/min at any pressure from 0 to 689.5 bar
PRESSURE RANGE 0.6895 to 689.5 bar
PRESSURE ACCURACY ± 0.5% of full scale at constant temperature
PRESSURE REPEATABILITY
d
± 0.5% of full scale within 48 hours at constant temperature
ZERO PRESSURE DRIFT ± 0.25% of full scale within 48 hours at constant temperature
PRESSURE DISPLAY RESOLUTION 6.895 kPa
AMBIENT TEMPERATURE RANGE 5 to 40°C
TEMPERATURE DRIFT ± 0.0.12% of full scale/°C
HUMIDITY 95% maximum
DEAD (HEADSPACE) VOLUME
e
1.30 ± 0.020 ml
POLLUTION DEGREE 2
INSTALLATION CATEGORY II
MAXIMUM ALTITUDE 2000 m
a. Underwriters Laboratories (UL) has certified all D Series Pumps with the exception of the 100 VAC versions.
b. Using water at 137.9 bar and a temperature controlled environment at 30°C.
c. The analog output is an optional accessory.
d. Pressure repeatability specification is based upon re-zeroing pressure transducer every 48 hours. Refer to sub-section
ZERO PRESS in Section 3 of the manual for re-zeroing procedure.
e. Volume in and above the piston seal, head clearance at automatic shutoff, and inlet and outlet ports to the fittings.
Factory Set
}