5 Using the product
28
2. Remove particle filter (1).
3. Remove the sealing plug (2) in the filter chamber from the
holder.
4. Close the gas path with the sealing plug.
5. Check that the sealing plug is fitted tightly. It should not yield at
all on being tugged gently.
CAUTION
Hot probe shaft! Risk of burns!
> Allow the probe shaft to cool down after a measurement,
before touching it!
> Only attach the silicone hose to the probe shaft once it has
cooled down!
6. Fit silicone hose onto the probe shaft of the flue gas probe. The
probe shaft openings must be closed.
Carry out the measurement
- The silicone hose must be clear (pressureless, no kinks).
1. Start measurement: [Start].
- Pressure zeroing.
2. Connect the silicone hose to the sampling point.