E.2.2 Replacing the probe module
1
Press the key on the top of the probe handle and
remove the probe module.
2
Fit a new probe module and engage it in place.
E.3 Regular care
E.3.1 Condensate trap
The fill level of the condensate trap can be read from the markings on the trap.
A warning message ( red flashing light) is displayed if the level in the condensate trap
reaches 90%.
Emptying the condensate trap
The condensate consists of a weak mix of acids. Avoid contact with the skin. Make
sure that the condensate does not run over the housing.
Condensate in gas path.
Damage tto mmeasuring ccells aand fflue ggas ppump.
Do not empty condensate trap while pump is operating.
1
Hold the measuring instrument so that the
condensate outlet points up.
2
Open condensate outlet in condensate trap: Pull
out approx. 5mm or until it will not go any further
(
).
3
Let the condensate run out into a sink (
).
4
Dab off drops at condensate outlet using a cloth.
5
Close the condensate outlet.
The condensate outlet must be fully closed (mark-
ing) otherwise incorrect measurements due to
inleaking air may result.
Placeholder:
Kondensatfalle_oeffnen.tif
Kondensatfalle_leeren.tif
E. Operation
E.2 Probes/Sensors
19