F.1.3 Alarm limits
A limit value can be set for ambient CO probes, leak detection probes and CO
2
probes.
A visual and optical alarm is triggered if a limit value is exceeded.
Calling up the function:
Inst’ settings Alarm limits .
Setting the alarm thresholds/ alarm signal :
Optional:
To reset all values to default values:
Default values .
1
Select the parameter.
Optional:
To reset the alarm threshold of the chosen parameter to the default value:
.
2
Set the values .
3
Repeat steps
1
and
2
for the other parameters/the alarm signal as required.
Saving settings:
OK Save input .
F.1.4 Start Keys edit
How the function keys are configured depends on the chosen function. Only the func-
tion keys in the start screen (shown when the measuring instrument is switched on) can
be assigned any function from the
Measurements menu.
The function keys are only active if the required probes are connected.
Calling up the function:
Inst’ settings Start Keys edit .
Assigning a function to the start keys:
1
Select function Press the function key that is to be assigned the chosen function.
2
Repeat step
1
for the other function keys as required.
Saving settings:
OK Save input .
OK
OKOK
OK
OKChange
Deflt
OK
OKOK