TESTO 330-1 - Editing and Managing Memory; Editing Location Details; Extras Memory Function; Instrument Diagnosis

TESTO 330-1
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Editing the location:
1
Select the location .
Options:
To delete the location: .
To change the location: .
Special function Extras memory
Calling uup aa ffunction:
Memory/Location .
- The available memory is displayed.
Options:
Print all data .
Delete all data .
Delete memory .
E.6 Instrument diagnosis
Important operating values and device data are displayed. A gas path check
(testo 330-2, -3 / testo 330-2LL) can be carried out. The status of the measuring cells
and device errors which have not yet been rectified can be viewed.
Calling up the function:
Inst’ diagnosis .
-oor-
.
Carrying out a gas path check (testo 330-2, -3 / testo 330-2LL)
1
Gas path check .
2
Place the black sealing cap on the tip of the flue gas probe.
- The pump flow is displayed. If the flow rate is 0,02l/min, the gas paths are not leak-
ing.
3
End the check: .
OK
OK
OK
OK
OK
OK
Extra
Change
Del
Edit