EasyManua.ls Logo

Thermo Scientific UltiMate - Page 19

Thermo Scientific UltiMate
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Process
OQ/PQ Operating Instructions M1393R0 Page 19 of 60
It is recommended that you perform the wash step separately and to use channel A or B
on such systems. Alternatively, the system can be upgraded with the new valve type (P/N
162297). Refer to Service Information #036 for more details.
If an oven is included in the UltiMate system, install the temperature probe (P/N 163962 )
in the oven compartment using the supplied screw and washer (FIGURE 3-4) and connect
the thermometer to a free COM port of the computer.
Built-in
Temperature
Sensor
Temperature
Probe
FIGURE 3-4 Installing the Temperature Probe
Note: Make certain that the low on battery indicator on the thermometer is not lit, because
a weak battery may interrupt the sequence if serial communication is broken.

Table of Contents

Related product manuals