5 Write Your Own Application
51
Command
Description
[UPPer]?
Queries the frequency range
LOWer?
POWer[:DC]
RANGe
AUTO {OFF|0|ON|1}
Switches the auto-ranging function on and off
AUTO?
Queries the auto-ranging function state
[:UPPer] {MINmum|MAXimum|
<numeric_valuje>[W]}
Sets the power range in W
[:UPPer]? [{MINimum|
MAXimum}]
Queries the power range
REFerence {MINimum|
MAXimum|DEFault|
<numeric_value>[W]}
Sets a delta reference value in W
REFerence? [{MINimum|
MAXimum|DEFault}]
Queries the delta reference value
STATe {OFF|0|ON|1}
Switches to delta mode
STATe?
Queries the delta mode state
UNIT {W|DBM}
Sets the power unit W or dBm
UNIT?
Queries the power unit
VOLTage[:DC]
RANGe
AUTO {OFF|0|ON|1}
Switches the auto-ranging function on and off
AUTO?
Queries the auto-ranging function state
[:UPPer] {MINmum|MAXimum|
<numeric_valuje>[V]}
Sets the current range in V
[:UPPer]? [{MINimum|
MAXimum}]
Queries the current range
REFerence {MINimum|
MAXimum|DEFault|
<numeric_value>[V]}
Sets a delta reference value in V
REFerence? [{MINimum|
MAXimum|DEFault}]
Queries the delta reference value
STATe {OFF|0|ON|1}
Switches to delta mode
STATe?
Queries the delta mode state
PEAKdetector
[:THReshold] {MINimum|
MAXimum|DEFault|
<numeric_value>
Sets the trigger level in % for the energy mode
[:THReshold]? [{MINimum|
MAXimum|DEFault}
Queries the trigger level setting