PM400
Command
Description
[:RESPonse] {MINimum|
MAXimum|DEFault|
<numeric_value>[A]}
[:RESPonse]?
[{MINimum|MAXimum|
DEFault}]
Queries the photodiode response value
:THERmopile
[:RESPonse] {MINimum|
MAXimum|DEFault|
<numeric_value>[V]}
Sets the thermopile response value in V/W
[:RESPonse]?
[{MINimum|MAXimum|
DEFault}]
Queries the thermopile response value
ENERgy
[:PYRO]
[:RESPonse] {MINimum|
MAXimum|DEFault|
<numeric_value>[V]}
Sets the pyro-detector response value in V/J
[:RESPonse]?
[{MINimum|MAXimum|
DEFault}]
Queries the pyro-detectro response value
CURRent[:DC]
RANGe
AUTO {OFF|0|ON|1}
Switches the auto-ranging function on and off
AUTO?
Queries the auto-ranging function state
[:UPPer] {MINmum|MAXimum|
<numeric_valuje>[A]}
Sets the current range in A
[:UPPer]? [{MINimum|
MAXimum}]
Queries the current range
REFerence {MINimum|
MAXimum|DEFault|
<numeric_value>[A]}
Sets a delta reference value in A
REFerence? [{MINimum|
MAXimum|DEFault}]
Queries the delta reference value
STATe {OFF|0|ON|1}
Switches to delta mode
STATe?
Queries the delta mode state
ENERgy
RANGe
[:UPPer] {MINmum|MAXimum|
<numeric_valuje>[J]}
Sets the energy range in J
[:UPPer]? [{MINimum|
MAXimum}]
Queries the energy range
REFerence {MINimum|
MAXimum|DEFault|
<numeric_value>[J]}
Sets a delta reference value in J
REFerence? [{MINimum|
MAXimum|DEFault}]
Queries the delta reference value
STATe {OFF|0|ON|1}
Switches to delta mode
STATe?
Queries the delta mode state
FREQuency
Range