◇23
4.2 Instruction of panel function interface and parameter
This section is mainly describe the function interface and relevant parameter in accordance with
the order of sofeware process and interface relevance.
Initial state introduction of the instrument
1. After starting up, the system enters into the last used setup interface before shutdown last
time.
2. The default set of the instrument is single step, AC withstanding voltage and default
parameters. As shown below.
3. The default cursor of the default interface is the interface switch. Other interface can be
chosen directly.
Default interface diagram of the instrument
Four function keys can be directly used to realize the interface switch, namely TEST, SETUP, SYSTEM
and FILE. The interface function will be introduced separately below.
Basic function operation of panel:
TEST: Instrument enters waiting status, the high voltage test starts.
SETUP: Change the previous interface of test programme, test item, and test parameter.
The change of test programme is finished in this interface.
SYSTEM: Enters the set interface related with test safety and the working mode.
FILE: Save and load the test programme, relating to the data storage.
: Cursor can move freely among each parameter.
F1~F5: Change the selected data by coordinating with the contents in soft key function display
zone.
NOTE: steps for system upgrade of the instrument:
1. Log-in the SystemSetup 2 and enter into the PASSWORD item.
2. Press LOCKSYSTEM and input the system lock password. The default password is
9310. Make sure the PASSWORD item is SYSTEM.
3. Turn off the instrument and then restart the instrument. Insert the USB-disk with
the upgrade file (USB must be FAT32 format and the capacity must be less than 4G.
The upgraded file must reside in a root directory).
4. Enter into the system password keyboard(F5) and input the start-up system upgrade
password. The password is 93102013.