ā—‡49
Data unit: mA
FUNC:SOURce:STEP:DC:WTIM To set /inquiry about the wait time for DCW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:DC:WTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:DC:WTIM?
--Data<sn>:
--Data<time>:
Data format: float
Data range: 0~999.9(0 is OFF)
Data accuracy: 0.1
Data unit: s
--Example:
Set the wait time for DCW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:DC:WTIM 1
Query message: FUNC:SOUR:STEP 1:DC:WTIM?
Return message: 1
FUNC:SOURce:STEP:DC:RAMP To set /inquiry about the RAMP state for DCW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:DC:RAMP<ON/OFF> or <1/0>
Query message:
FUNC:SOUR:STEP <sn>:DC:RAMP?
--Data<voltage>:
Data format: character
Data range: OFF(0), ON(1)
Data accuracy:
Data unit:
--Example:
Set the RAMP state for DCW test in STEP1 as ON.
Command message: FUNC:SOUR:STEP 1:DC:RAMP ON
Query message: FUNC:SOUR:STEP 1:DC:RAMP?
Return message: ON
5.3.5 Commands for IR Setup Function
NOTE: for the basic format, please refer tocommands for AC Setup function.
FUNC:SOURce:STEP:IR:VOLT To set /inquiry about the voltage for IR test
--Data<voltage>:
Data format: float
Data range: 50~1000
Data accuracy: 1