Query message:
FUNC:SOUR:STEP <sn>:AC:FREQ?
--Data<frequency>:
Data format: character
Data range: 50/60
Data accuracy:
Data unit: Hz
--Example:
Set the test frequency for ACW test in STEP1 as 50Hz.
Command message: FUNC:SOUR:STEP 1:AC:FREQ:50
Query message: FUNC:SOUR:STEP 1:AC:FREQ?
Return message: 50
FUNC:SOURce:STEP:AC:CH1 To set /inquiry about the milti-channel value
--Syntax:
Command message:
FUNCSOURSTEP 1ACCH1 <channel value>
Query message:
FUNCSOURSTEP 1ACCH1?
--Data<channel value>:
Data format: character
Data range: HIHG/LOW/OPEN
--Example:
Set the CH1 for ACW test in STEP1 as HIGH.
Command message: FUNCSOURSTEP 1ACCH1 HIGH
Query message: FUNCSOURSTEP 1ACCH1?
Return message: HIGH
NOTE: it shares the same method for other channel.
--Example:
Set the CH2 for ACW test in STEP1 as LOW.
Command message: FUNCSOURSTEP 1ACCH2 LOW
Query message: FUNCSOURSTEP 1ACCH2?
Return message: LOW
5.3.4 Commands for DC Setup Function
NOTE: for the basic format, please refer tocommands for AC Setup function.
FUNC:SOURce:STEP:DC:VOLT To set /inquiry about the voltage for DCW test
--Data<voltage>:
Data format: float
Data range: 50~6000
Data accuracy: 1
Data unit: V