ā46
--Data<time>:
Data format: float
Data range: 0~999.9 (0 is OFF)
Data accuracy: 0.1
Data unit: s
--Example:
Set the RISE time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:RTIM 1
Query message: FUNC:SOUR:STEP 1:AC:RTIM?
Return message: 1.000
FUNC:SOURce:STEP:AC:FTIM To set /inquiry about the FALL time for ACW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:FTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:FTIM?
--Example:
Set the FALL time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:FTIM 1
Query message: FUNC:SOUR:STEP 1:AC:FTIM?
Return message: 1
FUNC:SOURce:STEP:AC:ARC To set /inquiry about the ARC upper current for ACW test.
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:ARC<current>
Query message:
FUNC:SOUR:STEP <sn>:AC:ARC?
--Data<current>:
Data format: float
Data range: 0~20.0mA (0 is OFF) TH9320
Data accuracy: 0.1mA
Data unit: mA
--Example:
Set the ARC upper current for ACW test in STEP1 as 1mA.
Command message: FUNC:SOUR:STEP 1:AC:ARC 1
Query message: FUNC:SOUR:STEP 1:AC:ARC?
Return message: 1.000
FUNC:SOURce:STEP:AC:FREQ To set /inquiry about the test frequency for ACW test.
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:FREQ<frequency>