ā45
Command message: FUNC:SOUR:STEP 1:AC:UPPC 1
Query message: FUNC:SOUR:STEP 1:AC:UPPC?
Return message: 1.000
FUNC:SOURce:STEP:AC:LOWC To set /inquiry about the LOWER current for ACW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:LOWC<current>
Query message:
FUNC:SOUR:STEP <sn>:AC:LOWC?
--Data<current>:
Data format: float
Data range: 0~20.000 mA (0 is OFF) TH9320
0~10.000 mA (0 is OFF) TH9320
Data accuracy: 0.001mA
--Example:
Set the LOWER current for ACW test in STEP1 as 1mA.
Command message: FUNC:SOUR:STEP 1:AC:LOWC 1
Query message: FUNC:SOUR:STEP 1:AC:LOWC?
Return message: 1.000
FUNC:SOURce:STEP:AC:TTIM To set /inquiry about the TEST time for ACW test.
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:TTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:TTIM?
--Data<time>:
Data format: float
Data range: 0~999.9 (0 is OFF)
Data accuracy: 0.1
Data unit: s
--Example:
Set the TEST time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:TTIM 1
Query message: FUNC:SOUR:STEP 1:AC:TTIM?
Return message: 1
FUNC:SOURce:STEP:AC:RTIM To set /inquiry about the RISE time for ACW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:RTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:RTIM?