EasyManua.ls Logo

Tonghui TH9310 Series - Page 45

Tonghui TH9310 Series
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ā—‡45
Command message: FUNC:SOUR:STEP 1:AC:UPPC 1
Query message: FUNC:SOUR:STEP 1:AC:UPPC?
Return message: 1.000
FUNC:SOURce:STEP:AC:LOWC To set /inquiry about the LOWER current for ACW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:LOWC<current>
Query message:
FUNC:SOUR:STEP <sn>:AC:LOWC?
--Data<current>:
Data format: float
Data range: 0~20.000 mA (0 is OFF) TH9320
0~10.000 mA (0 is OFF) TH9320
Data accuracy: 0.001mA
--Example:
Set the LOWER current for ACW test in STEP1 as 1mA.
Command message: FUNC:SOUR:STEP 1:AC:LOWC 1
Query message: FUNC:SOUR:STEP 1:AC:LOWC?
Return message: 1.000
FUNC:SOURce:STEP:AC:TTIM To set /inquiry about the TEST time for ACW test.
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:TTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:TTIM?
--Data<time>:
Data format: float
Data range: 0~999.9 (0 is OFF)
Data accuracy: 0.1
Data unit: s
--Example:
Set the TEST time for ACW test in STEP1 as 1s.
Command message: FUNC:SOUR:STEP 1:AC:TTIM 1
Query message: FUNC:SOUR:STEP 1:AC:TTIM?
Return message: 1
FUNC:SOURce:STEP:AC:RTIM To set /inquiry about the RISE time for ACW test
--Syntax:
Command message:
FUNC:SOUR:STEP <sn>:AC:RTIM<time>
Query message:
FUNC:SOUR:STEP <sn>:AC:RTIM?

Table of Contents