◇50
Data unit: V
FUNC:SOURce:STEP:IR:UPPC To set /inquiry about the UPPER resistance for IR test
--Data< resistance>:
Data format: float
Data range: 0~1E4 (0 is OFF) MΩ
Data accuracy: 1.0 MΩ
Data unit: MΩ
FUNC:SOURce:STEP:IR:LOWC To set /inquiry about the LOWER resistance for IR test
--Data<resistance>:
Data format: float
Data range: 0.1~1E4 MΩ
Data accuracy: 0.1 MΩ
Data unit: MΩ
FUNC:SOURce:STEP:IR:TTIM To set /inquiry about the TEST time for IR test
--Data<time>:
Data format: float
Data range: 0~999.9 (0 is OFF)
Data accuracy: 0.1
Data unit: s
FUNC:SOURce:STEP:IR:RTIM To set /inquiry about the RISE time for IR test
--Data<time>:
Data format: float
Data range: 0~999.9 (0 is OFF)
Data accuracy: 0.1
Data unit: s
FUNC:SOURce:STEP:IR:FTIM To set /inquiry about the FALL time for IR test
--Data<time>:
Data format: float
Data range: 0~999.9 (0 is OFF)
Data accuracy: 0.1
Data unit: s
FUNC:SOURce:STEP:IR:RANG To set /inquiry about the range for IR test
--Data<range>:
Data format: integer
Data range:0~5 (0 is AUTO, 1 is 10mA, 2 is 2mA, 3 is 200uA, 4 is 20uA, 5 is 2uA)