◇56
SYST:OFFS?
--Data< ON/OFF/GET>:
Data format: character
Data range: 0(OFF), 1(ON), GET(get the reset value)
--Example:
Set the OFFS as ON,
Command message: SYST:OFFS ON or :SYST: OFFS 1
--Return message
Query message: SYST: OFFS?
If the state for OFFS is 1,
1 is returned.
--Gets the current value
Command message: SYST: OFFS GET
SYSTem:FAIL To set /inquiry about the state for AFTR FAIL
--Syntax:
Command message:
SYST:FAIL <0(STOP) /1(CONT) /2 (REST) /3( NEXT) >
Query message:
SYST:FAIL?
--Data< STOP/CONT/REST/NEXT>:
Data format: character
Data range: 0~3
--Example:
Set the AFTR FAIL as ON,
Command message: SYST:FAIL 0
--Return message
Query message: SYST: FAIL?
If the state for AFTR FAIL is 0,
0 is returned.
SYSTem:RESet To set all sets to default state.
--Syntax:
Command message:
Command message: SYST:RES
SYSTem:SMODE To set /inquiry about the working mode
--Syntax:
Command message:
SYST: SMOD <0(NORMAL)/1(REPEAT)
Query message:
SYST: SMOD?
--Data< working mode >:
Data format: integer