EasyManua.ls Logo

TPS 90-FL - User Manual

TPS 90-FL
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Congratulations !
The 90-FL is complete water quality logger in a single portable unit. It combines Dissolved
Oxygen, Conductivity, TDS, pH and Temperature.
Despite its impressive list of features, the 90-FL is a breeze to operate. This manual has been
designed to help you get started, and also contains some handy application tips. If at any stage you
require assistance, please contact either your local TPS representative or the TPS factory in
Brisbane.
The manual is divided into the following sections:
1. Table of Contents
Each major section of the handbook is clearly listed. Sub-sections have also been included to
enable you to find the information you need at a glance.
2. Introduction
The introduction has a diagram and explanation of the display and controls of the 90-FL. It
also contains a full listing of all of the items that you should have received with unit. Please
take the time to read this section, as it explains some of items that are mentioned in subsequent
sections.
3. Main Section
The main section of the handbook provides complete details of the 90-FL, including operating
modes, calibration, troubleshooting, specifications, and warranty terms.
4. Appendices
Appendices containing background information and application notes are provided at the back
of this manual.
TPS Pty Ltd
4 Jamberoo Street
Springwood, Brisbane,
Australia, 4127
Phone : (07) 32 900 400
International : 61 7 32 900 400
Fax : (07) 3808 4871
International : 61 7 3808 4871
E-mail : tps@tps.com.au
Web Site : www.tps.com.au
90-FL
Dissolved Oxygen,
Conductivity, TDS, pH,
Temperature Logger
Date : 01-Feb-2001
Author : MS
Version : 6.0

Table of Contents

Questions and Answers

  • S
    Steve LynchSep 12, 2025
    How to clean a blackened silver anode on TPS 90-FL Measuring Instruments?
    • P
      Phyllis WarrenSep 12, 2025
      If the silver anode on your TPS Measuring Instruments meter is blackened, it means the electrode has been exposed to sulphides or other chemical poisoning. Remove the pressure compensation diaphragm and membrane, then soak in 5% Ammonia solution for 10 minutes. If cleaning is unsuccessful, return the sensor to the TPS factory for cleaning and service.
  • K
    Kathryn MoranSep 9, 2025
    What causes 1 point calibration failures on TPS 90-FL?
    • J
      James FisherSep 9, 2025
      A 1-point calibration failure (pH asymmetry is greater than ±1.00 pH) on your TPS Measuring Instruments meter can be caused by a blocked reference junction or contaminated reference electrolyte. Clean the reference junction as per instructions supplied with the pH or reference electrode. Flush with distilled water and replace the electrolyte.
  • R
    robertwilsonSep 8, 2025
    Why do the readings drift on my TPS 90-FL?
    • J
      Johnny WilliamsSep 8, 2025
      If the readings on your TPS Measuring Instruments meter are drifting, it could be due to a build-up of dirt or oily material on the electrode plates of the sensor. Clean the sensor.
  • R
    ryan97Aug 22, 2025
    Why won't my TPS Measuring Instruments meter turn on?
    • R
      Richard LeeAug 22, 2025
      If your TPS Measuring Instruments meter won't turn on, the battery may be exhausted. Try recharging the battery for approximately 18 hours. If it still fails to turn on, check the charger. If the charger is working, replace the battery.
  • K
    Krystal ThomasSep 6, 2025
    Why does my TPS 90-FL attempt span calibration instead of zero calibration?
    • Z
      Zachary SchneiderSep 6, 2025
      If your TPS Measuring Instruments meter attempts span calibration instead of zero calibration, the sensor has a zero error. Thoroughly rinse the sensor in distilled water and allow it to completely dry in air before attempting zero calibration.
  • A
    Antonio BonillaSep 4, 2025
    Why does standard calibration fail with a low k factor on TPS 90-FL?
    • B
      Brandon ShermanSep 5, 2025
      A standard calibration failure with a k factor greater than 25% below the nominal value on your TPS Measuring Instruments meter can be due to: * Inaccurate standard solution; replace the standard solution. * A build-up of conductive material, such as salt, on the sensor; clean the sensor. * A faulty sensor. * A faulty instrument.
  • E
    Elizabeth ReillySep 3, 2025
    What causes standard calibration failures with high k factor on TPS 90-FL?
    • E
      Eric MuellerSep 3, 2025
      A standard calibration failure with a k factor greater than 25% above the nominal value on your TPS Measuring Instruments meter can be caused by: * The sensor not being immersed correctly; immerse sensor correctly as per diagrams. * A build-up of dirt or oily material on electrode plates; clean sensor as per the instructions. * A worn-off platinum-black coating; sensor requires replatinization or return to the factory for replatinization. * Inaccurate standard solution; replace standard solution. * A faulty sensor. * A faulty instrument.
  • A
    Ashley MorganSep 2, 2025
    What causes 2 point calibration failures on TPS 90-FL?
    • M
      Melissa GarrisonSep 2, 2025
      A 2-point calibration failure (pH slope is less than 85.0%) on your TPS Measuring Instruments meter can be caused by: * Incorrectly set pH buffers. For automatic pH buffer recognition, ensure that you are using buffers that match the selected buffer set. Otherwise, ensure that the buffer value is entered correctly at pH calibration. * A pH glass bulb that is not clean; clean the sensor surface or glass bulb as per instructions supplied with the electrode. * An aged electrode; attempt rejuvenation as per instructions supplied with the electrode. If unsuccessful, replace the electrode. * A damp connector; dry it in a warm place. * Inaccurate pH buffers; replace the buffers.
  • B
    Brian HuangSep 1, 2025
    Why does my TPS 90-FL fail to calibrate even with a new sensor?
    • D
      David SanchezSep 1, 2025
      If your TPS Measuring Instruments meter fails to calibrate even with a new sensor, it may be due to calibration settings being outside of allowable limits because of a previous failed calibration. Initialize the unit.
  • W
    williamwelchAug 31, 2025
    Why are the readings low or near zero on my TPS Measuring Instruments?
    • T
      Tammy AdamsAug 31, 2025
      Low or near-zero readings on your TPS Measuring Instruments meter can be due to several reasons: * The sensor may have a build-up of dirt or oily material on electrode plates; clean the sensor. * The sensor might not be immersed correctly; immerse the sensor as per the diagrams. * The sensor could be faulty. * The instrument itself might be faulty.

Summary

1. Introduction

1.1 90-FL Display and Controls

Details the physical layout and interactive elements of the 90-FL unit's front panel.

1.10 Specifications

Provides detailed technical specifications for the 90-FL instrument's various functions.

3. Dissolved Oxygen Mode

3.1 Selecting Dissolved Oxygen Mode

Guides on how to choose the desired units for Dissolved Oxygen readings.

3.2 Dissolved Oxygen Calibration

Details the process for calibrating the sensor using zero and span standards, including air and solution methods.

4. Conductivity Mode

4.2 Setting the Conductivity calibration standard

Explains how to input the standard solution value for Conductivity calibration.

4.3 Setting the Conductivity sensor k factor

Details how to select the correct k factor for different Conductivity sensors.

4.4 Conductivity Calibration

Outlines the steps for performing zero and standard calibrations for Conductivity measurements.

5. TDS Mode

5.2 Setting the TDS calibration standard

Explains how to input the standard solution value for TDS calibration.

5.3 Setting the TDS sensor k factor

Details how to select the correct k factor for different TDS sensors.

5.4 TDS Calibration

Outlines the steps for performing zero and standard calibrations for TDS measurements.

6. pH Mode

6.2 pH Calibration

Outlines the procedure for performing pH calibration using buffer solutions.

7. Temperature Mode

7.1 Temperature Calibration

Details the process for calibrating the temperature sensor for accurate readings.

8. Good Laboratory Practices (GLP)

9. Datalogging

9.2 Manually Recording Readings into the Logger

Details the process for manually saving measurements, including different A&B data input configurations.

9.3 Automatic Datalogging

Explains the different modes for automatic data logging: Rate per Day, Time of Day, and Sampling Period.

10. RS232 Port

12. Initialising the 90-FL

14. Battery Saver Function

16. Troubleshooting

16.1 General Errors

Lists common error messages, their causes, and remedies for general issues.

16.2 Dissolved Oxygen Troubleshooting

Provides solutions for specific problems encountered with Dissolved Oxygen measurements.

16.3 Conductivity; TDS Troubleshooting

Offers solutions for problems related to Conductivity and TDS sensor calibration and readings.

16.4 pH Troubleshooting

Provides solutions for specific issues encountered with pH measurements and calibration.

16.5 Temperature Troubleshooting

Lists troubleshooting steps for temperature measurement inaccuracies.

18. Warranty

TPS 90-FL Specifications

General IconGeneral
BrandTPS
Model90-FL
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals