Maintenance
•
To put the Atmos exhaust system into calibration mode, press On/Off button again and hold it down for at least 5
seconds. Release the key as soon as an acoustic signal sounds. The filter status indicator flashes yellow - start of
calibration. The Atmos Pure now calibrates the new filter with the resistance of the entire laser exhaust system. This
process takes about one minute.
•
The end of the calibration is confirmed with an acoustic signal and the exhaust system switches off.
8.3 Replace activated carbon and filter mats
1. Turn off the Atmos Pure.
2. Secure the Atmos against unexpected restarting by unplugging the power cord
from the power outlet.
3. Open the caps on both sides of the suction unit.
4. Remove the lid and li down the fine filter box of the Atmos.
5. Loosen all 4 screws of the filter frame and li it out with the grid.
6. Remove the filter mat.