30 Chapter 4
DEFAULT
Backwards compatible with Series 4000
/
4100 flow meters.
Returns the values for sample rate, calibration gas/gas mixture,
standard
/
volumetric flow units, display update rate, display mode,
analog zero, and full-scale scaling factors to the factory default
settings. This command also clears both the begin- and end-trigger
values. The default values for the Series 53XX
/
52XX operating
parameters are listed in Appendix B.
SN
Backwards compatible with Series 4000
/
4100 flow meters.
Returns the serial number of the flow meter in ASCII. The serial
number is an alpha-numeric string terminated by a <CR><LF>. The
string can be a maximum of 16 characters in length plus the
terminating <CR><LF>.
Example: 53101816001
MN
Backwards compatible with Series 4000
/
4100 flow meters.
Returns the model number of the flow meter in ASCII. The model
number is an alpha-numeric string terminated by a <CR><LF>. The
string can be a maximum of 12 characters in length plus the
terminating <CR><LF>.
Example: 531001
REV
Backwards compatible with Series 4000
/
4100 flow meters.
Returns the internal firmware revision of the flow meter in ASCII. The
revision is an alpha-numeric string terminated by a <CR><LF>. The
string can be a maximum of three characters in length plus the
terminating <CR><LF>.
Example: 1.3