EasyManuals Logo

TSI Instruments 5200 Series User Manual

TSI Instruments 5200 Series
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Command Set 31
HREV
Returns the internal hardware revision of the flow meter in ASCII. The
revision is an alpha-numeric string terminated by a <CR><LF>. The
string can be a maximum of three characters in length plus the
terminating <CR><LF>.
Example: A
DATE
Backwards compatible with Series 4000
/
4100 flow meters
Returns the date of the last calibration in ASCII. The format of the
string is month/day/year. The date is an alpha-numeric string
terminated by a <CR><LF>. The string can be a maximum of eight
characters in length plus the terminating <CR><LF>.
Example: 12/24/18
?
Backwards compatible with Series 4000
/
4100 flow meters
This is a ping command used to verify if the flow meter is
communicating.
The flow meter will return an acknowledge sequence of “OK
<CR><LF> as a response. The serial communications indicator will
flash once on the LCD display to indicate that the command was
received.
SUSTRxxxxxxxx
Backwards compatible with Series 4000
/
4100 flow meters.
Set a user-defined string to be stored in the meter.
SUSTR
Denotes Set user string command.
xxxxxxxx
x = alphanumeric character
Must enter eight characters following the command
NOTE: User string is not reset when the DEFAULT command is
used or during normal recalibration.

Other manuals for TSI Instruments 5200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments 5200 Series and is the answer not in the manual?

TSI Instruments 5200 Series Specifications

General IconGeneral
BrandTSI Instruments
Model5200 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals