Command Set 31
HREV
Returns the internal hardware revision of the flow meter in ASCII. The
revision is an alpha-numeric string terminated by a <CR><LF>. The
string can be a maximum of three characters in length plus the
terminating <CR><LF>.
Example: A
DATE
Backwards compatible with Series 4000
/
4100 flow meters
Returns the date of the last calibration in ASCII. The format of the
string is “month/day/year”. The date is an alpha-numeric string
terminated by a <CR><LF>. The string can be a maximum of eight
characters in length plus the terminating <CR><LF>.
Example: 12/24/18
?
Backwards compatible with Series 4000
/
4100 flow meters
This is a ping command used to verify if the flow meter is
communicating.
The flow meter will return an acknowledge sequence of “OK”
<CR><LF> as a response. The serial communications indicator will
flash once on the LCD display to indicate that the command was
received.
SUSTRxxxxxxxx
Backwards compatible with Series 4000
/
4100 flow meters.
Set a user-defined string to be stored in the meter.
Denotes Set user string command.
x = alphanumeric character
Must enter eight characters following the command
NOTE: User string is not reset when the DEFAULT command is
used or during normal recalibration.