32 Chapter 4
RUSTR
Backwards compatible with Series 4000
/
4100 flow meters.
Read the user-defined string to be stored in the meter.
RUSTR
Denotes read user string command.
Returns the 8 character user string.
SALIASxxxxxxxxxxxxxxxx
Set a user-defined string to be stored as the alias of the flow meter.
SUSTR
Denotes Set user string command.
xxxxxxxxxxxxxxxx
x = alphanumeric character (up to 16).
Must be up to 16 characters. String can be shorter.
NOTE: User string is not reset when the DEFAULT command is
used or during normal recalibration.
RALIAS
Read the user-defined string to be stored as the alias of the flow
meter.
RALIAS
Denotes read user string command.
Returns up to 16-character alias string.
BREAK
This command will stop the sending of data via the D commands
(DmFTPnnnn or DmFTPHLnnnn). When received, the meter will
stop sending data back via these commands. This command will also
stop and active volume commands.
Use this command to halt a data or volume command prior to
completion.