Using Serial Data Commands C-21
UAGenerate Accumulator Report Record
(Report Record A)
Enables and disables reporting of Record A in unpolled operation.
UAx
where:
x = 1 Enables A Record
x = 0 Disables A Record
UA will echo current setting 0 or 1.
See Record A in the “Unpolled Record Formats” section of this chapter.
UBGenerate Side Scatter Accumulator Report
Record (Report Record B)
Enables and disables reporting of Record B in unpolled operation.
UBx
where:
x = 1 Enables B Record
x = 0 Disables B Record
UB will echo current setting 0 or 1.
See Record B in the “Unpolled Record Formats” section of this chapter.
UCGenerate Correlated (Paired) Report Record
(Report Record C)
Enables and disables reporting of Record C in unpolled operation.
UCx
where:
x = 1 Enables C Record
x = 0 Disables C Record
UC will echo current setting 0 or 1.
See Record C in the “Unpolled Record Formats” section of this chapter.