C-22 Model 3321 Aerodynamic Particle Sizer
Spectrometer
UDGenerate Aerodynamic Data Report Record
(Report Record D)
Enables and disables reporting of Record D in unpolled operation.
UDx
where:
x = 1 Enables D Record
x = 0 Disables D Record
UD will echo current setting 0 or 1.
See Record D in the “Unpolled Record Formats” section of this chapter.
USGenerate Side Scatter Data Report Record
(Report Record S)
Enables and disables reporting of Record S in unpolled operation.
USx
where:
x = 1 Enables S Record
x = 0 Disables S Record
US will echo current setting 0 or 1.
See Record S in the “Unpolled Record Formats” section of this chapter.
UYGenerate Auxiliary Report Record (Report
Record Y)
Enables and disables reporting of Record Y in unpolled operation.
UYx
where:
x = 1 Enables Y Record
x = 0 Disables Y Record
UY will echo current setting 0 or 1.
See Record Y in the “Unpolled Record Formats” section of this chapter.