Using Serial Data Commands C-27
Auxiliary Data Record (Y)
CS,Y,bpress,tflow,sflow,a0,a1,d0,d1,d2,lpower,lcur,spumpv,tpumpv,itemp,
btemp,dtemp,Vop
CS = Checksum
Y = Auxiliary Data Record
bpress = barometric inlet pressure (average over sample time)
tflow = total flow (average over sample time)
sflow = sheath flow (average over sample time)
a0 = Analog input voltage 0(I/O connector pin 7 ref to pin 15
(average over sample time)
a1 = Analog input voltage 1(I/O connector pin 8 ref to pin 15
(average over sample time)
d0 = Digital input level of I/O connector pin 1 ref to GND (pins 9,
10)
d1 = Digital input level of I/O connector pin 2 ref to GND (pins 9,
10)
d2 = Digital input level of I/O connector pin 3 ref to GND (pins 9,
10)
lpower = laser power (% of maximum power)
lcur = laser current (ma)
spumpv = sheath pump voltage
tpumpv = total pump voltage
itemp = inlet temperature in degrees C (ex. 25.5)
btemp = internal box temperature in degrees C (ex. 31.5)
dtemp = detector temperature for APD and optics in degrees C (ex:
25.5)
Vop = APD operating voltage in volts
How t o I n put Com m a n ds and T r o u bleshoot
the Re s u l t s
Use the following information as a guide to inputting software commands
and for troubleshooting possible problems.
Input Guidelines
Input all alpha characters as capital letters (SMZ, not smz).
Separate parameters with commas, not spaces.
If you are in a command string, use the <Backspace> key to back up
and make changes. Do not use <arrow> keys.
At the end of a command string, press <Enter> to complete the string.