8.4. Diagnostics limits
Select "Diagnostics" -› "Diag. limits", and the following
display will appear:
Ctmp max =90 /OFF
Ptmp max =100.0 /ON
Ptmp min =0.0 /ON
Mlev min =10 /ON
R.Lev min =100.0 /ON
Mstab max =20 /ON
Rstab max =2 /ON
Eldr abs =80.0 /ON
VCOdr abs =50 /ON
The display shows the error limits for each dia-
gnostics measurement, and its effect on the current
signal (ON/OFF). To edit these limits press "Edit/Save"
and enter a new error limit, or change the ON/OFF
setting with the up/down arrow buttons.
Even if the effect of a limit on the current signal is
blocked by setting it OFF, an error message will still
appear in the status line of the main display, and error
data will also be stored in the error table.