Sensor
Measurement
Measuring range...... 0–40 % TS (if > 40 % TS, please
consult with Valmet Automation)
Repeatability ........... ± 0.01 % TS
Sensitivity ................ 0.001 % TS
Filtering ................... 1–99 s
Process conditions
pH-range ................. 2.5–11.5
Temperature............ +0–100 °C (+32–212°F)
Pressure:
- Minimum
recommend. .......... > 1.5 bar (> 22 psi), no free air
- Maximum .............. 16 bar / 100 bar (232 / 1450 psi)
Vibration .................. max. 20 m/s2, 10–200 Hz
Environmental conditions
Temperature............ -10…+70 °C (14…+158°F), protect
from direct heat sources
Housing class .......... IP66 (NEMA 4X)
ATEX ...................... Certificate No. VTT 12 ATEX 058X
II 3G Ex nR IIC T6 Gc
EN 60079-0:2010. EN 60079-15:2010
NOTE: Atex certificate does not
apply to sensor models with
flushing option
CRN ........................ 0F17146.5
Connections
Operating voltage,
communication ........ cable from TCU
Materials
Wetted parts ............ AISI 316, AISI 316L, Ceramic
Seals....................... EPDM
Operating unit TCU
Connections
Cable to sensor ....... 10 m (33 ft); option: 30 m (98 ft)
Operating voltage .... 90–260 VAC / 0.1 A
Connections to mill systems
- analog outputs ....... 2 current outputs (4–20 mA):
total solids content + process
temperature/conductivity (selectable)
- HART® ................. 12–36 VDC
- binary inputs .......... 2 x 12–28 VDC/10 mA, isolated
Connections to PC (start-up, service)
DTM for
FieldCare™ ............. HART®
PC connection
(service) .................. RS-232
Profibus PA
Environmental conditions
Temperature ............ -10...55°C (+14…131°F)
Housing class .......... IP65 (NEMA 4)
Dimensions & weight
w x h x d .................. 232 x 282 x 133 mm
(9 1/8" x 11 1/8" x 5 1/4")
Weight ..................... 2 kg (4.5 lbs)
Conductivity limits and sensor weights: