EasyManua.ls Logo

Vatech PHT-30CSS - Chapter 3. CT Image Evaluation; Spatial Resolution(MTF) Evaluation; Spatial Resolution (MTF) Standards; Phantom Specification

Vatech PHT-30CSS
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
100 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
Chapter 3. CT Image Evaluation
1. Spatial Resolution(MTF) evaluation
1.1. Spatial Resolution (MTF) standards
Parameter LSL (Lower Specification Limit) USL (Upper Specification Limit)
MTF 10% <lp/mm> 1.00 3.50
1.2. Phantom Specification
QUART: DVT Phantom

Table of Contents

Other manuals for Vatech PHT-30CSS

Related product manuals