Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
112 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
9.2. Test Method
1. Complete the collimator setup and acquire the image in the stopped state.
2. Run the C:\VCaptureSW\Util\EzEval_Phantom\EzEval.exe program.
3. Input the threshold 75%, and check the upper and lower.
4. Input the threshold 95%, and check the left and right.
5. Check whether the Collimator Alignment Result is satisfied with the standard.
9.3. Test Result
Parameter LSL Measurement value USL Pass/Fail
Threshold 95%
Left 1 2
Right 1 2
Threshold 75%
Upper 1 10
Lower 1 10