Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
157 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
3. CEPH Collimator Alignment
3.1. CEPH Collimator alignment
1. Run the C:\VCaptureSW\Acquisition\CEPH\WideCEPH\Norm\VACAP.exe file.
2. Send [SPM_CEPH], [SPM_HSON], [SPM_FREQ_0160], [SPM_COLM_0005], [SPM_CAMS],
[SPM_RET_], and [SPM_ERDY] commands to the Command window to configure the CEPH
sensor and collimator.
3. Set 90kvp, 10mA in VACAP, send the [SPM_SSB_] command to the Command window, and
click the Capture button to acquire the image.
4. Use the Collimator control command to find a value that satisfies the standard below.
Command Function
[SPM_CAST_XXXXX] CEPH mode collimator movement control
5. Run EzEval.exe, choose EzEvaluation for the tap menu on the top left and choose collimator
viewer from the tap menu on the top right.
6. Set the Threshold value concerning the following standard, and check 0010.raw to check if it
satisfies the collimator alignment standard.
7. The collimator alignment standard is as follows.
TOP
BOTTOM
RIGHT
LEFT

Table of Contents

Other manuals for Vatech PHT-30CSS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PHT-30CSS and is the answer not in the manual?

Vatech PHT-30CSS Specifications

General IconGeneral
BrandVatech
ModelPHT-30CSS
CategoryMedical Equipment
LanguageEnglish

Related product manuals