Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
87 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
5.2. CT Collimator Alignment with EzAlign
1. Select CBCT and click the Start button: The system will move into its capturing position and
Collimator auto-align will be started.
2. Remove the Bite Block.
3. When you see the message HOLD THE BUTTON as marked in the white box below, press
and hold the exposure switch until RELEASE THE BUTTON message appears.
4. If the system displays the HOLD THE BUTTON message again instead of the succeeding
message, repeat step 3.
5. If the CBCT Auto Alignment is completed successfully, the pop-up message which notifies
the completion of the alignment displays on the screen. Then, click the OK button to finish.
6. If the Error message appears, follow the guide. Then proceed the steps 1~5.