2.2. Check the sensor horizontal level
1. Place the digital level at the position as shown in the picture above, and check the horizontal
level. (Standard: ± 0.1° or less)
2. If the sensor level does not meet the standard, loosen the 4 point bolts in the red circle, adjust it
to meet the standard, and then refasten it.
2.3. Equipment Start Position Setting
1. As shown in the image above, set the Half value to [SPM_ACHF_XXXX] so that the guide line of
the rotator and the vertical are aligned in a straight line.
2. Send the same values for CT half value [SPM_ACHF_XXXX] in the PANO half
[SPM_HFST_XXXX] and the CEPH half [SPM_CPHF_XXXX].