Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
122 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
Collimator Alignment standard value
Collimator Min pixels Max pixels
Threshold
70%
Left 2 2
Right 1 2
Top 1 10
Lower 1 10
If Collimator Alignment was performed and changed values, you must
perform Collimator Speed Calibration.
Collimator Align can also be done using the EzAlign program provided
by the manufacturer.