10. Complete the CEPH Calibration, and choose the CEPH Processing button, as shown in [8].
11. When the Mode Confirm button is chosen, Level 1~3 images are created.
* Level 1~3: Level 3 has the highest contrast due to the difference in contrast.
12. Choose the O mark on the far right of , which is an image of Level 3 among the created
images.
13. When the Simulation button is chosen, an image of O marked in red, as shown in [9], is created.
14. Choose O marked in red and click the Apply button.
15. When a pop-up window is created, as shown in [10], click the OK button.
16. As shown in [11], check that the time of the modified date of Acquisition/WideCEPH /Norm or
Fast/UseParam/ZeusCEPH is the same as the time when the work is completed.