Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
202 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
031 - This error occurs while moving the
origin of the right side of the 4-axis
collimator.
Motor 4 Collimator right
axis error
032
- This error occurs while moving the
origin of the 4-axis collimator Up.
Motor 4 Collimator up axis
error
033 - This is an error that occurs while
moving the 4-axis collimator down the
origin.
Motor 4 Collimator down
axis error
034
- This error occurs while moving the 2-
axis collimator UP/Down origin.
Motor 2 Collimator
up/down axis error
035 - This error occurs while moving the
origin of the Left/Right 2-axis collimator.
Motor 2 Collimator left/right
axis error
036 - This error occurs while moving the
origin of the Left/Right 1-axis collimator.
Motor 1 Collimator left/right
axis error
037
- This error occurs during Generator
Tilting.
Motor generator tilting axis
error
039 - This error occurs while moving the
origin of the X-axis motor.
Motor X-axis error
040
- This error occurs while moving the
origin of the Y-axis motor.
Motor Y-axis error
041 - This error occurs while the sensor X-
axis motor origin is moving.
Motor Sensor X-axis error
003 ---- [ Exposure Switch related error ]
000 - Unknown error
060 - This error indicates that the exposure
switch is pressed when the device is
turned on.
EXP Push error
061 - This is an error that stops X-Ray
exposure by releasing the exposure
switch.
004 ---- [ Other errors ]
000 - Unknown error
100
- Headrest error. In PANO mode, the
headrest must be removed.
Headband support error
101 - Hand frame error. The hand frame is
open. Equipment stops running.
Hand frame support error
102
- This error occurs when CAN
communication is not responding.
CAN communication error
103 - This error occurs when the column
position sensor is not responding.
Column position sensor
error
104 - This error occurs when the column is
not automatically adjusted.
Column auto control error

Table of Contents

Other manuals for Vatech PHT-30CSS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PHT-30CSS and is the answer not in the manual?

Vatech PHT-30CSS Specifications

General IconGeneral
BrandVatech
ModelPHT-30CSS
CategoryMedical Equipment
LanguageEnglish

Related product manuals