78
11 Supplement
VEGAPULS 63 • Modbus and Levelmaster protocol
41364-EN-170711
Additional output parameter - Electronics temperature
Output of the values
Ʋ Analogue Via the current output
Ʋ Digital Via the digital output signal (depending on the electron-
ics version)
Range
-40 … +85 °C (-40 … +185 °F)
Resolution < 0.1 K
Accuracy ±3K
Voltage supply
Operating voltage
8 … 30 V DC
Power consumption < 500 mW
Reverse voltage protection Integrated
Potential connections and electrical separating measures in the instrument
Electronics
Notnon-oating
Ground terminal Galvanicallyconnectedwiththemetalprocesstting
Galvanic separation between electronics and metal housing parts
Ʋ Reference voltage 500 V AC
Electrical protective measures
Protection rating
Housing material Version Protection acc. to
IEC 60529
Protection acc. to NEMA
Plastic Single chamber
IP 66/IP 67 Type4X
Double chamber IP 66/IP 67 Type4X
Aluminium Single chamber IP 66/IP 68 (0.2 bar)
IP 68 (1 bar)
Type 6P
-
Double chamber
IP 66/IP 68 (0.2 bar)
IP 68 (1 bar)
Type 6P
-
Stainless steel (electro-
polished)
Single chamber
IP 66/IP 68 (0.2 bar) Type 6P
Stainless steel (precision
casting)
Single chamber IP 66/IP 68 (0.2 bar)
IP 68 (1 bar)
Type 6P
-
Double chamber
IP 66/IP 68 (0.2 bar)
IP 68 (1 bar)
Type 6P
-
Connection of the feeding power supply
unit
Networks of overvoltage category III
Altitude above sea level
Ʋ by default bis 2000 m (6562 ft)
Ʋ with connected overvoltage protection up to 5000 m (16404 ft)