G4 Multi Channel Weighing Instrument
Exit calibration.
To exit from calibration, press key ↑ twice.
A question, 'Save calib?', is displayed.
Answer Yes (key ↵ ) to keep the new settings
Save calib?
in the instrument memory, or
No Esc. Yes
answer No (key – ) to exit calibration without saving
(all changes will be rejected).
In both cases the instrument switches to sub menu 'Calibration'.
Answer Esc. (key ↑ ) to continue the calibration for Scale 1.
Now it is possible to perform calibration of any remaining scales,
or to return to normal weight display by pressing key ↑ several times.