Operating instructions, Quick installation
Zeroing of gross weight
When the scale installation is calibrated, a zeroing is performed in order to make
the gross weight zero when the scale is unloaded. If mechanical equipment is added
later, this will affect the weight display for the scale, and a new zeroing of the gross
weight becomes necessary.
NOTE! The scale must be unloaded when zeroing is performed.
Access ‘Set Zero' for any calibration type.
Navigate to sub menu 'Calibration' and select the
desired scale, in this example Scale 1.
The sequence is described on page 25.
Press key ↵ .
Use key + to scroll the parameter list until
line 'Scale 1 Scale 1 Calib' is displayed.
Press key ↵ .
Use key + to scroll the parameter list until
parameter '1:Set Zero' is displayed.
Press key ↵ .
The Set-up Code or the Operator Code may be
needed to continue.
Display of live gross weight gives the possibility
to check the load on the scale.
Press key ↵ .
The value +000000.00 kg is proposed for '1:Set Zero'.
Press key ↵ for 1 second to accept 0.00 kg.
The zero setting is now finished.
To exit the zero setting, press key ↑ twice.
A question, 'Save calib?', is displayed.
Answer Esc. (key ↑ ) to continue the zero setting of
Scale 1.
Answer Yes (key ↵ ) to keep the new zero setting, or
answer No (key – ) to exit calibration without saving (all changes will be rejected).
In both cases the instrument switches to menu 'Calibration'.
Press key ↑ several times to return to weight display.
Calibration
Scale 1
Scale 1
Scale 1 Calib
1:Set Zero
0.00 kg
Gross Weight:
xxx.xx
1:Set Zero
+000000.00 kg
1:Set Zero
0.00 kg
Save calib?
No Esc. Yes
36