| Description of the device
22 | CO2 Incubator Basic VWR
Chapter 4
Door switch
Figure 6 – Door switch
A door switch [1] is installed in the upper edge of the work space opening. If the door switch is
activated by opening the glass door, the gas supply and the heating of the work space are
interrupted and the display text will flash.
If the door remains open for more than 30 seconds, a short acoustic alarm sounds. If the door
remains open for more than 10 minutes, the acoustic alarm sounds continuously.
The outer door can only be closed after the glass door has been latched properly.