VWR CO2 Incubator Basic | 61
Maintenance
Inspections and checks
To ensure the operativeness and the operational safety of the device, the functions and device
components listed below must be checked at different intervals.
Daily check:
Gas supply of the CO
2
supply system
Annual inspection:
Tightness of the glass door seal
Permeability of the pressure compensation opening with insert
Functional check of the operating panel and of the device control
Electrical safety check in accordance with the relevant national regulations (e.g. VBG 4)
NOTE!
Functional check:
If safety devices were removed or disabled for inspections, the device must not be operated
before the safety devices have been reinstalled and checked for their correct function!
Service intervals
During normal operation, the following service routine must be performed:
Weekly service:
Refill the work space water tray with fresh processed water
3-month service:
Run auto-start routine
Perform temperature and CO
2
calibration