VWR CO2 Incubator Basic | 57
Cleaning and disinfection |
Chapter 9
90 °C decontamination routine
The entire program run of the decontamination routine takes approx. 25 hours.
CAUTION
Hot surface:
The surfaces of the exterior door handle, work space, particularly the glass door armatures and the
interior sheet of the outer door, are heated during the decontamination routine. During the routine
run or immediately after completion of the run, always wear safety gloves when touching these
surfaces; observe the warning indicator at the operating panel!
1. After the cleaning, reinstall the shelf system components into the work space.
2. Fill the water tray with 350 ml of processed water.
3. Turn the device on using the power switch.
4. Activate decontamination routine.
5. After the decontamination routine has been completed, remove the remaining water using a
sterile cloth.
6. Turn the device off or restart the device operation using auto-start routine.
NOTE!
Duration of the auto-start routine:
When the device is cold and when the ambient temperature is low, the auto-start routine may take
up to 10 hours.
90 °C decontamination routine procedure
Figure 18 – 90 °C decontamination routine procedure
The routine is divided into four phases. Each individual phase or several phases can be canceled
(i.e. skipped). If the operating step “Cancel decontamination routine” is executed, the routine
moves to the next program phase. To cancel the routine completely, the operating step must be
executed repeatedly until the remaining run time display shows the value 0. When the glass door is
opened, this operating step cancels the routine completely.