EasyManua.ls Logo

VWR CO2 Incubator Basic - Chapter 12 Technical Data

VWR CO2 Incubator Basic
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VWR CO2 Incubator Basic | 73
Technical data
TECHNICAL SPECIFICATIONS
DESCRIPTION UNITS: VALUE:
MECHANICAL
External dimensions
Width x Depth x Height
mm
in
637 x 867 x 766
25.1 x 34.1 x 30.2
Interior dimensions
Width x Depth x Height
mm
in
470 x 607 x 530
18.5 x 23.9 x 20.9
Chamber volume
l approx. 151
cu.ft approx. 5.3
Shelves
Width x Depth
mm
in
423 x 455
16.7 x 17.9
Standard quantity Piece 3
Maximum quantity Piece 10
Maximum surface load Piece 10 / insertion shelf
Maximum device overall load kg
ib
30
66.1
Weight, without accessories kg
ib
70 (stainless steel)
154.3 (stainless steel)
THERMAL
Ambient temperature range
°C +18...33
Temperature control range °C RT + 3 ... 55
Temperature deviation at 37°C, time °C ± 0.1
Temperature deviation, spatial (DIN 12880, Part 2) at 37 °C °C ± 0.6
Duration of the auto-start routine, to 37 °C ambient temperature
20 °C
h 5 ... 10
Temperature recovery time, at 37 °C, door open 30 seconds
(to 98% of initial value)
min < 10
Heat transfer to environment:
at 37 °C
during 90 °C decontamination
kWh/h
kWh/h
0.06
0.26

Table of Contents