| Technical data
74 | CO2 Incubator Basic VWR
Chapter 12
HUMIDITY
Water quality demineralized / distilled or
autoclaved or completely
deionized / distilled or
autoclaved.
Liquid quantity:
Incubation operation
90°C decontamination operation
l
ml
max. 3.0 / min 1.2
300
Constant humidity at 37 °C % rH approx. 93
Humidity recovery time, at 93% rH, door open 30 s
(to 95% of initial value)
min
min
approx. 20 (stainless steel)
CO
2
GAS SUPLLY SYSTEM
Gas purity % min. 99.5 or medical quality
Prepressure bar min. 0.8 - max. 1
Measuring and control range % vol. 0... 20
Control deviation, temporal % vol. ± 0.1
Recovery time, at 5%, door open 30 seconds
(to 98% of initial value)
min < 8
CO
2
MEASURING CELL
Accuracy (absolute) % CO
2
± 0.3
ELECTRICAL SYSTEM
Rated voltage V 1/N/PE 120 V, AC
Rated frequency Hz 50/60
Interference suppression (DIN VDE 0875) Interference level N
Type of protection (DIN 40 050) IP 20
Protection class l
Overvoltage category (IEC 1010, EN 61010) ll
Pollution severity (IEC 1010, EN 61010) 2
Rated current A 5.3 (120 V, AC)
On-site fusing:
Fuse
Circuit breaker
T 16 A
G 16
Rated input kW 0.64 (120 VAC)
EMC class B
OTHERS
Sound pressure level (DIN 45 635, Part 1) dB(A) < 50
Relative humidity of environment % rH max. 80
Location elevation m NN max. 2000
TECHNICAL SPECIFICATIONS
DESCRIPTION UNITS: VALUE: